ventilator portable adult 20X10-1/2X8"operates in control assist
adult; 20 in. x 10-1/2 in. x 8 in.; operates in control, ASSIST-Control, and continuous positive airway pressure modes w/or without positive end expiratory pressure; peep will be controllable from 0-20 cm of water; equipped w/mixing valve, airway gage; audible and visual alarm to warn low pressure; manual cycled capable; for use w/monitor, oxygen nsn 6515-01-130-1379 and field anesthesia apparatus nsn 6515-01-003-4133; for support kit nsn 6545-01-267-5565
An item designed to provide ventilation to a patient in transport and/or during institutional care. It may be in a transport case and provide time cycled, pressure limited, piston driven constant flow, be microprocessor controlled, with three modes, control, assist-control, synchronous intermittent mandatory ventilation providing sensitivity control, rechargeable battery, hoses, tubing, connectors, patient valve, and other related features.
Indicates there is no data in the hmirs and the nsn is in a fsc not generally suspected of containing hazardous materials.
Item does not contain precious metal.
The item does not have a nuclear hardened feature or any other critical feature such as tolerance, fit restriction or application.
HMIC: Hazardous Material Indicator Code. A one position code that identifies a hazardous item.
PMIC: Precious Metal Indicator Code. A one position code which identifies items that have precious metals as part of their content. precious metals are those metals generally considered to be uncommon, highly valuable, and relatively superior in certain properties such as resistance to corrosion and electrical conductivity.
ESD: Electrostatic Discharge. Indicates if an item is susceptible to electrostatic discharge or electromagnetic interference damage. electrostatic discharge damage occurs when an accumulation of static electricity generated by the relative motion or separation of materials is released to another item by direct contact. electromagnetic interference damage occurs when an item comes into proximity with an electrostatic or magnetic field.
ENAC: Enviromental Attribute Code. Identifies items with environmentally preferred characteristics.
CRITL: Criticality Indicator Code. Indicates an item is technically critical by tolerance, fit, application, nuclear hardness properties, or other characteristics.